This series of temperature sensors is used to measure the temperature of gas andliquid media. The maximum measurable pressure is 1.0 MPa.
The thermometer is equipped with a welded protective sheath as a whole, and is provided with quick-installation clamps and unions and other sanitary process connections. It is mainly applied in thesanitary industry. The electrical connection adopts common junction boxes and sanitary junction boxes.
The food and beverage industry,pharmaceutical industry, surfactant production, biotechnology and life science engineering, dairy plants,breweries
Thermoresistance temperature range: -200C to +600C. Itcan be integrated with a transmitter and is programmable andcalibratable.
The process connection and probe are made of stainless steel.
Electrical connection fomm:Junction box
Both the materials and surface treatment quality comply with the standards of hygienic design.
Output signal
The thermometer provides Pt100 signal direct output or 4-20mA transmitter output.
Direct output type
| Temperature range | Absence of cervical :-50℃~150℃ With neck tube : -50℃~250℃ |
| Measuring element | Thermistors |
| Wiring method | ■Two-wire system, ■3-wire system, ■4-wire system |
| Tolerance values of measuring elements | ■Grade B ■Grade A |
| Output format | Resistance signal |
| Electrical connection | ■Ordinary junction box ■Explosion-proof junction box ■Display junction box |
| Explosion-proof protection (optional) | EXdIICT4 isolated explosion-proof |
4...20mA Transmitter Output Type
| Temperature range | Absence of cervical :-50℃~150℃ With neck tube : -50℃~250℃ |
| Measuring element | Pt100 thermal resistance |
| Tolerance values of measuring elements | ■Grade B ■Grade A |
| asic configuration | 0 - 150℃ (default) |
| Output format | 4-20mA, 2-wire system |
| Load RA | RA ≤ (UB - 9V) / 0.023A |
| Load impact | ±0.05%/100 Ω |
| power supply | DC 12 - 32V |
| Power protection | Reverse protection |
| Impact of power supply variations | ±0.025%/V |
| Electromagnetic Compatibility (EMC) | Immunity and burst pulses |
| Electrical connection | ■Ordinary junction box ■Explosion-proof junction box ■Display junction box |
| Explosion-proof protection (optional) | In EXdIICT4 |
Environmental conditions
| Ambient temperature and storage temperature | -40°C to 85°C Direct output type -40°C to 120°C |
| Degree of protection | IP 65 is suitable for junction boxes. |
| Response time | < 10 seconds (applicable to sensors with a diameter of 6mm) |
| Materials | stainless steel |
| Seismic resistance | 3g (IEC 60751, standard) |